# ICS Advisory (ICSMA-18-179-01)
## Medtronic MyCareLink Patient Monitor
Original release date: June 28, 2018
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-179-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-179-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-179-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 6.4**
* **Vendor:** Medtronic
* **Equipment:** MyCareLink Patient Monitor
* **Vulnerabilities:** Use of Hard-coded Password, Exposed Dangerous Method or Function
## 2\. RISK EVALUATION
If exploited, these vulnerabilities may allow privileged access to the
monitor's operating system. However, physical access to the MyCareLink monitor
is required. Additionally, these vulnerabilities may allow a MyCareLink
monitor, when operated within close physical proximity of an implantable
cardiac device, to read and write arbitrary memory values of that device.
## 3\. TECHNICAL DETAILS
### 3.1 AFFECTED PRODUCTS
The following MyCareLink Monitors are affected:
* 24950 MyCareLink Monitor, all versions,
* 24952 MyCareLink Monitor, all versions.
### 3.2 VULNERABILITY OVERVIEW
**3.2.1 [USE OF HARD-CODED PASSWORD
CWE-259](https://cwe.mitre.org/data/definitions/259.html)**
The affected product contains a hard-coded operating system password. An
attacker with physical access can remove the case of the device, connect to
the debug port, and use the password to gain privileged access to the
operating system
[CVE-2018-8870](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8870)
has been assigned to this vulnerability. A CVSS v3 base score of 6.4 has been
assigned; the CVSS vector string is
([AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:H/PR:N/UI:N/S:U/C:H/I:H/A:H)).
**3.2.2 [EXPOSED DANGEROUS METHOD OR FUNCTION
CWE-749](https://cwe.mitre.org/data/definitions/749.html)**
The affected product contains debug code meant to test the functionality of
the monitor's communication interfaces, including the interface between the
monitor and implantable cardiac device. An attacker with physical access to
the device can apply the other vulnerabilities within this advisory to access
this debug functionality.
This debug functionality provides the ability to read and write arbitrary
memory values to implantable cardiac devices via inductive or short range
wireless protocols. An attacker with close physical proximity to a target
implantable cardiac device can use this debug functionality.
[CVE-2018-8868](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-8868)
has been assigned to this vulnerability. A CVSS v3 base score of 6.2 has been
assigned; the CVSS vector string is
([AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L)).
### 3.3 BACKGROUND
* **CRITICAL INFRASTRUCTURE SECTORS:** Healthcare and Public Health
* **COUNTRIES/AREAS DEPLOYED:** Worldwide
* **COMPANY HEADQUARTERS LOCATION:** Dublin, Ireland
### 3.4 RESEARCHER
Peter Morgan of Clever Security reported these vulnerabilities to NCCIC.
## 4\. MITIGATIONS
Medtronic will release several rolling over-the-air product updates that will
mitigate the vulnerabilities described within this advisory. These updates
will be applied to devices automatically as part of standard, reoccurring
update processes. In addition, Medtronic has increased security monitoring of
affected devices and related infrastructure.
Medtronic recommends users take additional defensive measures to minimize the
risk of exploitation of these vulnerabilities. Specifically, users should:
* Maintain good physical controls over the home monitor as the best mitigation to these vulnerabilities.
* Only use home monitors obtained directly from their healthcare provider or a Medtronic representative to ensure integrity of the system.
* Report any concerning behavior regarding their home monitor to their healthcare provider or a Medtronic representative.
Medtronic has released additional patient focused information, at the
following location:
<https://www.medtronic.com/security>
NCCIC recommends users take defensive measures to minimize the risk of
exploitation of this vulnerability. Specifically, users should:
* Minimize network exposure for all control system devices and/or systems, and ensure that they are [not accessible from the Internet](/ics/alerts/ICS-ALERT-10-301-01).
* Locate control system networks and remote devices behind firewalls, and isolate them from the business network.
* When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
NCCIC reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
NCCIC also provides a section for [control systems security recommended
practices](/ics/content/recommended-practices) on the ICS-CERT web page.
Several recommended practices are available for reading and download,
including [Improving Industrial Control Systems Cybersecurity with Defense-in-
Depth Strategies](/sites/default/files/recommended_practices/NCCIC_ICS-
CERT_Defense_in_Depth_2016_S508C.pdf).
Additional mitigation guidance and recommended practices are publicly
available on the [ICS-CERT website](/ics/) in the Technical Information Paper,
[ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation
Strategies](/ics/tips/ICS-TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to NCCIC for
tracking and correlation against other incidents.
No known public exploits specifically target these vulnerabilities. These
vulnerabilities are not exploitable remotely. High skill level is needed to
exploit.