# ICS Advisory (ICSMA-18-156-01)
## Philips' IntelliVue Patient and Avalon Fetal Monitors
Original release date: June 05, 2018
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-156-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-156-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-156-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 8.3**
* **Vendor:** Philips
* **Equipment:** IntelliVue Patient Monitors, Avalon Fetal/Maternal Monitors
* **Vulnerabilities:** Improper Authentication, Information Exposure, Stack-based Buffer Overflow
## 2\. RISK EVALUATION
Successful exploitation may allow an attacker to read/write memory, and/or
induce a denial of service through a system restart, thus potentially leading
to a delay in diagnosis and treatment of patients.
## 3\. TECHNICAL DETAILS
### 3.1 AFFECTED PRODUCTS
The following IntelliVue Patient Monitors versions are affected:
* IntelliVue Patient Monitors MP Series (includingMP2/X2/MP30/MP50/MP70/NP90/MX700/800) Rev B-M, and
* IntelliVue Patient Monitors MX (MX400-550) Rev J-M and (X3/MX100 for Rev M only).
The following Avalon Fetal/Maternal Monitors versions are affected:
* Avalon Fetal/Maternal Monitors FM20/FM30/FM40/FM50 with software Revisions F.0, G.0 and J.3
### 3.2 VULNERABILITY OVERVIEW
### 3.2.1 [IMPROPER AUTHENTICATION
CWE-287](https://cwe.mitre.org/data/definitions/287.html)
The vulnerability allows an unauthenticated attacker to access memory ("write-
what-where") from an attacker-chosen device address within the same subnet.
[CVE-2018-10597](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-10597)
has been assigned to this vulnerability. A CVSS v3 base score of 8.3 has been
calculated; the CVSS vector string is
([AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:H/A:H)).
### 3.2.2 [INFORMATION EXPOSURE
CWE-200](https://cwe.mitre.org/data/definitions/200.html)
The vulnerability allows an unauthenticated attacker to read memory from an
attacker-chosen device address within the same subnet.
[CVE-2018-10599](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-10599)
has been assigned to this vulnerability. A CVSS v3 base score of 6.4 has been
calculated; the CVSS vector string is
([AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:L](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:U/C:H/I:L/A:L)).
### 3.2.3 [STACK-BASED BUFFER OVERFLOW
CWE-121](https://cwe.mitre.org/data/definitions/121.html)
The vulnerability exposes an "echo" service, in which an attacker-sent buffer
to an attacker-chosen device address within the same subnet is copied to the
stack with no boundary checks, hence resulting in stack overflow.
[CVE-2018-10601](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-10601)
has been assigned to this vulnerability. A CVSS v3 base score of 8.2 has been
calculated; the CVSS vector string is
([AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:A/AC:H/PR:N/UI:N/S:C/C:H/I:L/A:H)).
### 3.3 BACKGROUND
* **CRITICAL INFRASTRUCTURE SECTORS:** Healthcare and Public Health
* **COUNTRIES/AREAS DEPLOYED:** Worldwide
* **COMPANY HEADQUARTERS LOCATION:** Netherlands
### 3.4 RESEARCHER
Oran Avraham of Medigate, in coordination with Philips, reported these
vulnerabilities to NCCIC.
## 4\. MITIGATIONS
Philips will provide a remediation patch for specific supported versions, as
well as an upgrade path for all versions. Philips will communicate service
options to all affected install-base users.
Philips recommends users obtain associated field change and service bulletin
information from Philips by accessing their InCenter account at this location:
<http://incenter.medical.philips.com> (link is external).
Please see the Philips product security website for the latest public security
information on this matter and for other Philips products:
<https://www.philips.com/productsecurity> (link is external).
Philips provides the following mitigations for these vulnerabilities:
* In order for users of affected devices to mitigate exposure to these vulnerabilities, Philips recommends following the device's labeling, including Instructions for Use and Service Guide(s), which provide compensating controls to mitigate these vulnerabilities. Philips will provide remediation for IntelliVue software Revisions J-M and Avalon software Revisions G.0 and J.3 in 2018 in the form of a patch. For users with earlier IntelliVue software revisions, Philips provides an upgrade path for a currently-supported software revision. For upgrade options, users should contact their Philips sales representative.
* For IntelliVue Monitors - To mitigate these vulnerabilities; Philips recommends users follow the instructions for use (Security for Clinical Networks Guide) physical and logical security. Additionally, Philips recommends users update to Revision K.2 or newer software.
* For Avalon Fetal Monitors Release G.0 and Release J.3 - Philips recommends users follow the installation and service manual (Data Privacy and Network Security Requirements).
* For Avalon Fetal Monitors Release F.0 - Philips recommends users follow the instructions as documented in the Rev J.3 Service Guide Data Privacy and Network Security Requirements section.
* Users with questions regarding their specific IntelliVue and Avalon Fetal Monitor installations are advised by Philips to contact their local Philips service support team or their regional service support.
* Philips also recommends the following:
* The Philips Security for Clinical Networks guide and the IntelliVue Clinical Networks Configuration Guide, which provide instructions for the use of physical security access controls.
* The Avalon Revision J.3 Service Guide Data Privacy and Network Security Requirements section provides instructions for use of physical and logical security access controls.
* The Philips Security for Clinical Networks Guide and the IntelliVue Clinical Networks Configuration Guide, which provide instructions for use of logical security access controls. There is no clinical requirement for these devices to communicate outside of the Philips clinical network.
* Locate all medical devices and remote devices behind firewalls and isolate them from the business network following the recommendations documented in the Philips Security for Clinical Networks Guide, and the IntelliVue Clinical Networks Configuration Guide.
NCCIC recommends that users take defensive measures to minimize the risk of
exploitation of these vulnerabilities. Specifically, users should:
* Minimize network exposure for all control system devices and/or systems, and ensure that they are [not accessible from the Internet](/ics/alerts/ICS-ALERT-10-301-01).
* Locate all medical devices and remote devices behind firewalls, and isolate them from the business network.
* When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
NCCIC reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
NCCIC also provides a section for [control systems security recommended
practices](/ics/content/recommended-practices) on the ICS-CERT web page. NCCIC
reminds organizations to perform proper impact analysis and risk assessment
prior to deploying defensive measures.
Additional mitigation guidance and recommended practices are publicly
available on the [ICS-CERT website](/ics/) in the Technical Information Paper,
[ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation
Strategies](/ics/tips/ICS-TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to NCCIC for
tracking and correlation against other incidents.
No known public exploits specifically target these vulnerabilities. These
vulnerabilities are exploitable from within the same local device subnet. High
skill level is needed to exploit.