# ICS Advisory (ICSMA-18-128-01)
## Silex Technology SX-500/SD-320AN or GE Healthcare MobileLink (Update B)
Original release date: May 08, 2018 | Last revised: June 05, 2018
[Print Document](javascript:window.print\(\);)
[Tweet](https://twitter.com/share?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-128-01)
[Like Me](https://www.facebook.com/sharer.php?u=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-128-01)
[Share](http://www.addthis.com/bookmark.php?url=https%3A%2F%2Fus-
cert.cisa.gov%2Fics%2Fadvisories%2FICSMA-18-128-01)
### Legal Notice
All information products included in [https://us-cert.gov/ics](/ics) are
provided "as is" for informational purposes only. The Department of Homeland
Security (DHS) does not provide any warranties of any kind regarding any
information contained within. DHS does not endorse any commercial product or
service, referenced in this product or otherwise. Further dissemination of
this product is governed by the Traffic Light Protocol (TLP) marking in the
header. For more information about TLP, see [https://www.us-
cert.gov/tlp/](/tlp/).
* * *
## 1\. EXECUTIVE SUMMARY
* **CVSS v3 7.4**
* **ATTENTION** : Exploitable Remotely / Low skill level to exploit / Public exploits are available
* **Vendors** : Silex Technology, GE Healthcare
* **Equipment** : SX-500, SD-320AN, MobileLink
* **Vulnerabilities** : Improper Authentication, OS Command Injection
## 2\. UPDATE INFORMATION
This updated advisory is a follow-up to the updated advisory titled
ICSMA-18-128-01 Silex Technology SX-500/SD-320AN or GE Healthcare MobileLink
(Update A) that was published May 31, 2018, on the NCCIC/ICS-CERT website.
## 3\. RISK EVALUATION
Successful exploitation of these vulnerabilities could allow modification of
system settings and remote code execution.
## 4\. TECHNICAL DETAILS
### 4.1 AFFECTED PRODUCTS
The following products from Silex Technology are affected. Some are not
affected by both vulnerabilities (see Mitigations Section):
* GEH-500 Version 1.54 and prior (integrated into GE MobileLink),
* SX-500 All Versions (end-of-life 2011),
* GEH-SD-320AN Version GEH-1.1 and prior (integrated into GE MobileLink), and
* SD-320AN Version 2.01 and prior (end-of-life Nov 2017).
The following models of GE MAC Resting ECG analysis system may use the
vulnerable MobileLink technology:
* MAC 3500,
* MAC 5000 (end-of-life 2012),
* MAC 5500, and
* MAC 5500 HD.
### 4.2 VULNERABILITY OVERVIEW
### 4.2.1 [IMPROPER AUTHENTICATION
CWE-287](https://cwe.mitre.org/data/definitions/287.html)
Authentication is not verified when making certain POST requests, which may
allow attackers to modify system settings.
[CVE-2018-6020](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-6020)
has been assigned to this vulnerability. A CVSS v3 base score of 6.5 has been
calculated; the CVSS vector string is
([AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:L](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:L)).
### 4.2.2 [IMPROPER NEUTRALIZATION OF SPECIAL ELEMENTS USED IN AN OS COMMAND
('OS COMMAND INJECTION')
CWE-78](https://cwe.mitre.org/data/definitions/78.html)
A system call parameter is not properly sanitized, which may allow remote code
execution.
[CVE-2018-6021](http://web.nvd.nist.gov/view/vuln/detail?vulnId=CVE-2018-6021)
has been assigned to this vulnerability. A CVSS v3 base score of 7.4 has been
calculated; the CVSS vector string is
([AV:N/AC:L/PR:L/UI:N/S:C/C:L/I:L/A:L](https://www.first.org/cvss/calculator/3.0#CVSS:3.0/AV:N/AC:L/PR:L/UI:N/S:C/C:L/I:L/A:L)).
### 4.3 BACKGROUND
* **Critical Infrastructure Sectors:** Healthcare and Public Health
* **Countries/Areas Deployed:** Worldwide
* **Company Headquarters Location:** United States
### 4.4 RESEARCHER
Eric Evenchick of Atredis Partners reported these vulnerabilities to Silex and
GE and tested pre-release firmware and other mitigations confirming they
resolved the vulnerabilities.
## 5\. MITIGATIONS
Silex Technologies and GE Healthcare recommend the following mitigations:
CVE-2018-6020 (GE MobileLink/SX-500) - Enable the "update" account within the
web interface, which is not enabled by default. Set the secondary password for
the "update" account to prevent unauthenticated changes to the device
configuration.
**\--------- Begin Update B Part 1 of 1 --------**
CVE-2018-6021 (GE MobileLink/GEH-SD-320AN) - Silex Technology and GE
Healthcare have produced an updated firmware image (Version 1.14) for the GEH-
SD-320AN, which is now available for download at the following location:
<http://silextechnology.com/geh320an/>
GE Healthcare has released a security notice that can be viewed at:
<http://www3.gehealthcare.com/en/support/security>
The firmware update (Version 2.0.3) for SD-320AN is separate from GEH-SD-320AN
and is available for download from Silex Technology at the following link:
<https://www.silextechnology.com/connectivity-solutions/device-
connectivity/sd-320an>
Note that this update does not pertain to the listed GEH device. Contact Silex
Technology for more information regarding download and application of this new
firmware.
**\--------- End Update B Part 1 of 1 --------**
NCCIC recommends that users take defensive measures to minimize the risk of
exploitation of these vulnerabilities. Specifically, users should:
* Minimize network exposure for all control system devices and/or systems, and ensure that they are [not accessible from the Internet](/ics/alerts/ICS-ALERT-10-301-01).
* Locate control system networks and remote devices behind firewalls, and isolate them from the business network.
* When remote access is required, use secure methods, such as Virtual Private Networks (VPNs), recognizing that VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize that VPN is only as secure as the connected devices.
NCCIC reminds organizations to perform proper impact analysis and risk
assessment prior to deploying defensive measures.
NCCIC also provides a section for [control systems security recommended
practices](/ics/content/recommended-practices) on the ICS-CERT web page.
Several recommended practices are available for reading and download,
including[ Improving Industrial Control Systems Cybersecurity with Defense-in-
Depth Strategies](/sites/default/files/recommended_practices/NCCIC_ICS-
CERT_Defense_in_Depth_2016_S508C.pdf).
Additional mitigation guidance and recommended practices are publicly
available on the [ICS-CERT website](/ics/) in the Technical Information Paper,
[ICS-TIP-12-146-01B--Targeted Cyber Intrusion Detection and Mitigation
Strategies](/ics/tips/ICS-TIP-12-146-01B).
Organizations observing any suspected malicious activity should follow their
established internal procedures and report their findings to NCCIC for
tracking and correlation against other incidents.
暂无评论